AVS47-IB SPECIFICATIONS

Command summary (firmware version 3R4)

NOTE 1: All commands listed below are sequential
NOTE 2: All responses are single-line
NOTE 3: "c" means command and "q" means query


COMMANDS FOR THE AVS-47B

REM local/remote c/q
INP input zero/measure/100 ohms calibrate c/q
MUX multiplexer channel c/q
RAN range c/q
EXC nominal excitation voltage c/q
DIS display selector c/q
REF deviation reference voltage (command only)
RFS reference source (query only)
MAG difference magnifier (query only)
ARN manual/autorange (by software) c/q
SDY stabilisation delay c/q
ZDR take reference from display (command only)
ADC make one A/D conversion (command only)
     ADC? ask results of the latest ADC as integer (q)
     RES? ask results as resistance (q)
AVEn record average on n A/D conversions (c)
     STP stop averaging immediately (c)
     AVE? ask for average resistance (q)
     MIN? ask for minimum (q)
     MAX? ask for maximum (q)
     STD? ask for standard deviation of resistance (q)
OVL? ask if overload during last A/D conversion or average (q)



COMMANDS FOR THE TS-530A

SPT set point (digital) (c/q)
PRO proportional gain (c/q)
PRO15 force error signal to zero (c)
ITC integrator time constant (c/q)
DTC derivator time constant (c/q)
BIA bias power (c/q)
POW heater power range (c/q)
SPV measure/ask set point voltage (c/q)
HTV measure/ask heater output voltage (c/q)
HTI measure/ask heater output current (c/q)
HTP measure and calculate/ask heater power (c/q)



DIGITAL FILTER COMMANDS

DFL digital filter length (c/q)
DFS start digital filter (c)
     STP stop digital filtering (c)
     DFR ask for a reading (q)



SCANNING COMMANDS

FCH set the first channel to be scanned (c/q)
LCH set the last channel to be scanned (c/q)
SCI set the scanning interval (c/q)
ETC enable simultaneous temp. control (c/q)
TCP temperature control period (c/q)
SCPn set channel parameters for channel n (c/q)
SDY stabilisation delay (c/q)
CNT average count (c/q)
EXC excitation for channel n (c/q)
RAN range for channel n (c/q)
SCNx start to scan (0:single cycle, 1:front panel started mode, 2:continuous) (c)
     STP stop scanning immediately (c)
SCRx Scan results command for channel x
     AVE?     ask for average
     MIN?      ask for minimum
     MAX?      ask for maximum
     STD?      ask for standard deviation



GENERAL AVS47-IB COMMANDS

TIM set/get real time (c/q)
DAY set/get real time date (c/q)
PRN printer on/off (c/q)
DSK ram-disk buffer on/off (c/q)
PBF print buffer file (c)
STP stop printing buffer file (c)
RBF start reading buffer file (c)
NXT give next line (q)
STP stop reading buffer file immediately (q)
BUF read buffer size and free space (q)
KIL delete the buffer file (c)
DLYx insert delay of x secods (c)
PONRST simulates power-on reset, does not wipe the buffer (c)



IEEE-488.2 COMMON COMMANDS AND QUERIES

*IDN? identification query (q)
*STB? read status byte query (q)
*SRE set/get service request enable mask (c/q)
*ESE set/get standard event status register enable mask (c/q)
*CLS clear status command (c)
*OPC operation complete command and query (c/q)
*RST device reset command (c)
*TST self-test query (includes also the TS-530A) (q)



MISCELLANEOUS COMMANDS

IBA IEEE-488 device address (c/q)
HDR include message headers in responses (c/q)
PBA picobus device address (default=1) (c/q)
PBD picobus delay factor (default set by factory) (c/q)
LRS measure and calculate appr. lead resistance (c/q)
SCAL self-calibration macro command and query (c/q)
FSM disable/enable front panel started measurement (c/q)
EWC Eeprom write count query (q)



SCANNING

Scanning is made using preselected excitation voltages for each sensor. Several A/D conversions can be averaged in order to get a better resolution. Switching between sensors is made in a sequence that prevents accidental overheating. Software autoranging can be used.
    The TS-530A temperature controller can be connected to sensor No. 0. Because the AVS-47B Resistance Bridge can measure only one channel at a time, the controller is set automatically to a state where its heater output remains constant during the measurement of other sensors. Active control is enabled for a selectable period (TCP) between each channel to be scanned. Simultaneous scanning and control is suitable only in systems where thermal conditions remain reasonably stable.


PRINTING AND BUFFERING

Results can be buffered in a 64kB memory together with time stamps. The buffered results can be either printed on paper or read via the GBPI bus. Printing/reading is always sequential, starting from the first record. Measurement results can also be printed in real time to an LPT: printer by using the front panel START/PRINT switch.


DIGITAL FILTERING

The digital filter is a simple running average of a pre-set number of readings. An automatic re-start follows changes in settings, like selecting a new resistance range (autoranging). During digital filtering no other commands are accepted except STP (stop) and DFR (give digital filter reading).


DIGITAL CALIBRATION

The digital self-calibration procedure measures gain factors and offsets on all seven excitation ranges and uses this information to correct the readings. This feature is convenient it the AVS-47B suffers from large changes in operating temperature. Calibration data is volatile and calibration must be repeated whenever the AVS47-IB power has been off.




MECHANICAL DATA

Size: 210mm(W), 160mm(D), 65mm(H)
Weight: 1 kg
Mains voltage range: 90-250V 50-60Hz.


Back to AVS47-IB description